TF: 1.844.ESC(372).8877

APEX PREMIER – HIGH PERFORMANCE FUME HOOD MONITOR

The Apex Premier™ utilizes state-of-the-art technology to provide the highest level of performance achievable in a fume hood monitor.

The heart of the Apex Premier is the smart sensor module which is responsible for accurate, repeatable and instantaneous measurement of fume hood face velocity.  Face velocity measurement is made by an ultra-sensitive micro-bridge mass airflow sensor with a response time of less than 3 milliseconds.

The analog signal provided by the micro-bridge is digitized by a high-resolution A/D converter, then further enhanced and factory calibrated over the operating velocity range by an automated test an calibration system.

Applications

All types of fumehoods including high performance low flow hoods, variable volume and constant volume.

Advantages

Microprocessor-based performance
Microbridge mass airflow sensor technology and advanced automated factory calibration
Real-time graphical display of face velocity
Internal menu driven setup and field calibration
Programmable mute button, output relays for Purge or Alarm, digital inputs for alarm or setback
Entire system settings viewable via one button
Password protected calibration configuration
Self-contained face velocity sensor

TECHNICAL DATA SHEET

Copyright © 2015 ESC. All rights reserved.